MODEL 2113F and 2114F CHARGE AMPLIFIERS

MODEL 2113F and 2114F CHARGE AMPLIFIERS

2113f

BOUCHE LABS MODEL 2113F and 2114F CHARGE AMPLIFIERS

BOUCHE LABS Model 2113F and 2114F Precision Charge Amplifiers are designed for use in a comparison system along with a Model 1000A Primary Vibration Standard Shaker. They can also be used for comparison shock calibrations. The Model 2113F is used to condition the output of the standard accelerometer built into the armature of the Model 1000A; the Model 2114F is used to condition the output of a piezoelectric test accelerometer. Each amplifier includes an individual power supply, and is designed for stability and precision necessary in a calibration system. An optional low pass filter is available for suppression of resonances during shock calibrations.



Specifications

Measurement Unit Rating
INPUT 12
MODE Precision gain single-ended charge converter
GROUND (Insulated from case) ohms 10M
TRANSFER +/-2
GAIN RANGES
(2113F) mV/g 10 000, 1 000, 100, 10, 1
(2114F) mV/g 1 000, 100, 10, 1, .1
ACCURACY (on any range) % +/-0.3
GAIN STABILITY
maximum (ref: line voltage, 105 to 125 Vac) % 0.1
maximum per 1000 pF change in source capacitance % 0.1
FREQUENCY RESPONSE, ref: 100 Hz
Filter Out: 10 Hz to 10 000 Hz % +/-0.5
at 5 Hz: % +0.2/-1.0
Filter in: 5 Hz to 15 kHz % +/-5
at 22 kHz, minimum dB -3
at 28 kHz, minimum dB -10
Terminal slope dB/Oct -18
AMPLITUDE LINEARITY (ref:best straight line) % +/-0.2
WARM-UP TIME min 1
HARMONIC DISTORTION, maximum % 0.2
TRANSIENT DISTORTION
maximum time delay of optional low-pass filter, any frequency 5 Hz to 15 kHz us 4
POWER 1/4-28, 10-32, etc.
LINE VOLTAGE Vac 105 to 125
CIRCUIT GROUND, Insulated from case, minimum ohm 10M
Characteristic Unit Rating
PHYSICAL
OVERALL DIMENSIONS in/cm 1.75 x 4.5 x 10/4.4 x 11.4 x 25.4
FINISH Anodized Aluminum
WEIGHT lb 3
CONNECTORS
Input 10-32 Microdot S-50 or equivalent
Output BNC, UG-1094/U or equivalent
ENVIRONMENTAL
OPERATING TEMPERATURE °F/°C +40 to +125/+5 to +52
STORAGE TEMPERATURE °F/°C -65 to +185/-54 to +85
RELATIVE HUMIDITY, maximum % 95
°C -54 to +85
RELATIVE HUMIDITY, maximum % 95
Accessories Supplied

Instruction Manual and Input and Output Mating Connectors.


Calibration

Reciprocity calibration of Model 2113F with 1000A unit is recommended.

BOUCHE LABS reserves the right to change these and any other specifications without notice.

Share by: